2 To edit the parameter you currently have selected, press
3 To confirm your choice, press the appropriate softkey.
4 To add a new frequency dependent offset table entry, press
5 To remove a frequency dependent offset table entry, use the
You can only edit the table name, you cannot delete it.
N O T E
6 When you have completed editing the frequency dependent offset table,
Ensure that the frequency points you use cover the frequency range of the signals you
N O T E
want to measure. If you measure a signal with a frequency outside the frequency range
defined in the frequency dependent offset table, the power meter uses the highest or
lowest frequency point in the frequency dependent offset table to calculate the offset.
E4419B Power Meter User's Guide
An offset in the range of 1% to 150% can be entered.
The following rules apply to naming frequency dependent offset tables:
• The name must consist of no more than 12 characters.
• All characters must be upper or lower case alphabetic characters or
numeric (0-9) or an underscore (_).
• No other characters are allowed.
• No spaces are allowed in the name.
power meter displays the data in a pop-up window. Modify this data (see
below) as desired.
• Use
or
to modify the character on which the cursor is currently
positioned.
• Use
or
to move to other characters.
prompted for the frequency and offset. The entry is sorted by frequency.
, and
keys to select the entry, press
frequency, the equivalent offset is also removed and vice versa.
press
to return to the "Offset Tbls" screen.
Done
Power Meter Operation
. The
Change
Insert
,
. If you delete the
Delete
2
. You are
,
61