Table 2-3 TTL Input Timing Diagram 1

Table 2-4 TTL Inputs Timing Diagram 2

E4418B Power Meter User's Guide
T4
Time for zero/cal operation to complete.
All timings based on 100 ms firmware polling.
Timing of zero/cal inputs for condition "00".
Inputs
Operation
Time
Description
T5
Maximum time between inputs going low.
T6
Minimum overlap of low inputs.
T5
T6
T7
Power Meter Operation
Zero: 10 s (8480 Series)
12 s (E-Series)
20 s (N8480 Series exclude
Option CFT)
8 s (N8480 Series with Option
CFT)
Cal: 6 s (8480 Series)
7 s (E-Series)
9 s (N8480 Series exclude
Option CFT)
6 s (N8480 Series with Option
CFT)
T8
Value
100 ms
200 ms
2
37