10 You must set the display to the type of measurement you require. This can
11 The measurement result is now displayed.
If the measurement frequency does not correspond directly to a frequency in
N O T E
the sensor calibration table, the power meter calculates the calibration factor
using linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor
calibration table, the power meter uses the highest or lowest frequency point
in the sensor calibration table to set the calibration factor.
The value of the calibration factors being used by the power meter to make a
measurement is displayed under the

Editing Sensor Calibration Tables

You can edit sensor calibration tables using the "Edit Cal" menu as shown in
Figure
To view the sensor calibration tables currently stored in the power meter,
press
displayed as shown in
E4419B Power Meter User's Guide
either be a direct channel measurement, a ratio measurement of both
channels or a difference measurement between both channels. Use the
following key presses:
Press
,
Input Select
measurement you require,
2-5.
,
,
Tables Sensor Cal Tables
Figure
. Select the appropriate softkey for the
,
,
,
, or
A B A/B B/A
or
A Cal Fac
. The "Sensr Tbls" screen is
2-5.
Power Meter Operation
, then
or
A-B
softkeys.
B Cal Fac
2
.
B-A
49