Table 5-5 Settling Time - Agilent Technologies E4419B User Manual

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

5
Specifications
For Agilent E-Series power sensors
In FAST mode, within the range –50 dBm to +17 dBm, for a 10 dB decreasing
power step, the settling time is
1
point, add 25 ms. Refer to the relevant power sensor manual for further
information.
For Agilent E- Series power sensors in normal and x2 speed modes, manual
filter, 10 dB decreasing power step:

Table 5-5 Settling Time

Number of Averages
Settling Time (s)
(Normal Mode)
Response Time (s)
(x2 mode)
174
When a decreasing power step crosses the power sensor's auto- range switch
1
2
4
0.07
0.12
0.21
0.04
0.07
0.12
8
16
32
64
0.4
1
1.8
3.3
0.21
0.4
1
1.8
128
256
512
6.5
13
27
3.4
6.8
14.2
E4419B Power Meter User's Guide
1024
57
32