11 The measurement result is now displayed.
If the measurement frequency does not correspond directly to a frequency in the sensor
N O T E
calibration table (if selected) and the frequency dependent offset table being used, the
power meter calculates the calibration factor and offset using linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor calibration table
or the frequency dependent offset table, the power meter uses the highest or lowest
frequency point in the appropriate table to set the calibration factor and offset.
The value of the calibration factors being used by the power meter to make a
measurement is displayed under the
E4419B Power Meter User's Guide
following key presses:
Press
,
Input Select
measurement you require,
. Select the appropriate softkey for the
,
,
,
, or
A B A/B B/A
or
A Cal Fac
B Cal Fac
Power Meter Operation
, then
or
A-B
B-A
softkeys.
2
.
59