Figure 6-5 Troubleshooting Guide - Ttl Inputs - Agilent Technologies E4416A Service Manual

Power meters
Hide thumbs Also See for E4416A:
Table of Contents

Advertisement

6
Troubleshooting
Investigate Power
Meter Failure

Figure 6-5 Troubleshooting Guide - TTL Inputs

122
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Start
Power Up
meter
POST
No
Pass?
Yes
Connect Sensor to Channel A
Connect Sensor to Channel B (if applicable)
Any Sensor
Errors?
No
Select
System/Inputs
More
Service
Self Test
Instrument Self Test
Self Test
Yes
Errors?
No
Select
Done
Zero/Cal
More
TTL Inputs On
Apply TTL logic 0 pulse to RJ45 connector pin 5
wrt pin 8. Apply TTL logic 1 pulse to RJ45 connector
pin 6 wrt pin 8. See User's Guide Ch. 2 for timing.
A
End
Yes
Investigate
Sensor Failure