Agilent Technologies E4419B User Manual page 183

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

–361
–362
–363
–410
–420
–430
–440
User's Guide
Parity error in program
The serial port receiver has detected a parity error and consequently, data integrity cannot be
guaranteed.
Framing error in program
The serial port receiver has detected a framing error and consequently, data integrity cannot be
guaranteed.
Input buffer overrun
The serial port receiver has been overrun and consequently, data has been lost.
Query INTERRUPTED
A command was received which sends data to the output buffer, but the output buffer
contained data from a previous command (the previous data is not overwritten). The output
buffer is cleared when power has been off, or after *RST (reset) command has been executed.
Query UNTERMINATED
The power meter was addressed to talk (that is, to send data over the interface) but a
command has not been received which sends data to the output buffer. For example, you may
have executed a CONFigure command (which does not generate data) and then attempted
to read data from the remote interface.
Query DEADLOCKED
A command was received which generates too much data to fit in the output buffer and the
input buffer is also full. Command execution continues but data is lost.
Query UNTERMINATED after indefinite response
The *IDN? command must be the last query command within a command string.
Error Messages