Agilent Technologies E4419B User Manual page 182

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

4
Error Messages
–310
–310
–310
–320
–330
–330
–330
–330
–330
–330
–330
–330
–350
160
System error;Sensor EEPROM < > data not found or unreadable
Where < > refers to the sensor data block covered, for example, Linearity, Temp - Comp
(temperature compensation).
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
System error;Option 001 Battery charger fault
The power meter is connected to an AC power source, the battery is not fully charged and it is
not charging.
System error;Sensors connected to both front and rear inputs.
You cannot connect two power sensors to the one channel input. In this instance, the power
meter detects power sensors connected to both its front and rear channel inputs.
Out of memory
The power meter required more memory than was available to run an internal operation.
Self-test Failed;
The –330, "Self-test Failed" errors indicate that you have a problem with your power meter.
Refer to
"Contacting Agilent Technologies"
faulty power meter.
Self-test Failed;Option 001 Battery requires replacement
The Option 001 battery is not charging to a satisfactory level and should be replaced.
Self-test Failed;RAM Battery Fault
Refer to
"RAM Battery"
on page 104 if you require a description of the battery test.
Self-test Failed;Calibrator Fault
Refer to
"Calibrator"
on page 105 if you require a description of the calibrator test.
Self-test Failed;ROM Check Failed
Refer to
"ROM Checksum"
test.
Self-test Failed;RAM Check Failed
Refer to
"RAM"
on page 104 if you require a description of the RAM test.
Self-test Failed;Display Assy. Fault
Refer to
"Display"
on page 105 if you require a description of the Display test.
Self-test Failed;Serial Interface Fault
Refer to
"Serial Interface"
Queue overflow
The error queue is full and another error has occurred which could not be recorded.
on page 109 for details of what to do with your
on page 104 if you require a description of the ROM Checksum
on page 104 if you require a description of this test.
E4419B Power Meter User's Guide