Configuring the Remote Interface

This section explains how to configure the power meter for remote control
using either the GPIB parallel interface or the RS232/422 serial interface. For
further information on using the power meter remotely, refer to the
Agilent EPM- 4418B/4419B Programming Guide.

GPIB

To select GPIB as the remote interface, press
Select Interface GPIB
GPIB Address
Each device on the GPIB (IEEE-488) interface must have a unique address.
You can set the power meter's address to any value between 0 and 30. The
address is set to 13 when the power meter is shipped from the factory.
The address is stored in nonvolatile memory and does not change when the
power is switched off or after a remote interface reset.
Your GPIB bus controller has its own address. Avoid using the bus controller's
address for any instrument on the interface bus.
To set the GPIB address from the front panel:
1 Press
2 To change this setting, press
3 To confirm your choice, press
To set the GPIB address from the remote interface, use the:
• SYSTem:COMMunicate:GPIB:ADDRess command.
86
,
,
Remote Interface Configure Interface GPIB
setting of the GPIB address is displayed under the
address in a pop-up window. Modify this address (see below) as desired.
• Use
or
to modify the digit on which the cursor is currently
positioned.
• Use
or
to move to other digits.
(IEEE 488).
,
GPIB Addr
. The power meter displays the
Enter
.
,
Remote Interface
,
. The current
softkey.
GPIB Addr
E4419B Power Meter User's Guide
,