Agilent Technologies E4419B User Manual page 158

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

3
Menu Reference
136
Serial
Press this softkey to access a menu which allows you to set baud rate,
word size, parity, number of stop bits, pacing and echo for the serial
interface.
Baud Rate
Press this key to set the serial interface baud rate. The default value is
9600. A pop-up window appears on the screen giving a range of values
(50, 75, 110, 150, 300, 1200, 1800, 2000, 2400, 3600, 4800, 7200, 9600,
19.2K, 38.4K, 57.6K, or 115.2K).
Word Size
Press this softkey to set the serial interface word length. A pop-up
window appears on the screen giving a selection of seven or eight.
Stop Bits
Press this softkey to set the serial interface number of stop bits. A
pop-up window appears on the screen giving a selection of one or two.
Parity
Press this softkey to enable/disable parity and set the parity type for
the serial interface. A pop-up window appears on the screen allowing
you to select from ODD, EVEN, ZERO, ONE or NONE.
Pacing
Press this softkey to access a menu which allows you to enable/disable
various hardware and software pacing options.
Tx Pacing
Press this softkey to toggle between enabling and disabling the
Xon/Xoff transmitter software handshake. When enabled, "Xon" is
highlighted, otherwise "None" is highlighted.
Rx Pacing
Press this softkey to toggle between enabling and disabling the
Xon/Xoff receiver software handshake. When enabled, "Xon" is
highlighted, otherwise "None" is highlighted.
RTS/CTS
Press this softkey to obtain a pop-up window which allows you to
select one of the following:
OFF - Sets the RTS signal line permanently low.
ON - Sets the RTS signal line permanently high.
E4419B Power Meter User's Guide

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents