Settling Time; Table 5-4 Settling Time - Agilent Technologies E4419B User Manual

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

Settling Time

0 to 99% settled readings over the GPIB.
For Agilent 8480 Series power sensors
Manual filter, 10 dB decreasing power step:

Table 5-4 Settling Time

Number of Averages
Settling Time (s)
(Normal Mode)
Response Time (s)
(x2 mode)
Auto filter, default resolution, 10 dB decreasing power step,
normal and x2 speed modes:
Normal Mode
Maximum dBm
150 ms
Typical
Settling
200 ms
Times
500 ms
6.6 s
Minimum dBm
E4419B Power Meter User's Guide
1
2
4
8
0.15
0.2
0.3
0.5
0.15
0.18
0.22
0.35
20 dB
Power
Sensor
10 dB
Dynamic
Range
10 dB
10 dB
16
32
64
1.1
1.9
3.4
0.55
1.1
1.9
x2 Mode
Maximum dBm
150 ms
Typical
Settling
180 ms
Times
350 ms
3.5 s
Minimum dBm
Specifications
128
256
512
1024
6.6
13
27
57
3.5
6.9
14.5
33
20 dB
Power
Sensor
10 dB
Dynamic
Range
10 dB
10 dB
5
173