Battery Option 001 Operational Characteristics - Agilent Technologies E4419B User Manual

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

5
Specifications

Battery Option 001 Operational Characteristics

The following information describes characteristic performance based at a
temperature of 25
product performance that is useful in the application of the product but is not
covered by the product warranty.
Typical Operating Time
Up to two hours with LED backlight on; up to three hours with LED backlight
off.
Charge Time
Less than two hours to charge fully from an empty state: 50 minutes charging
enables one hour of operation with LED backlight ON; 35 minutes charging
enables one hour of operation with the LED backlight OFF. Power meter is
operational while it is charging.
Service Life
To 70% of initial capacity at 25 °C: approximately 450 charge/discharge cycles.
Chemistry
Nickel Metal Hydride.
Weight
1 kg.