Table 3-1 Combination Of Sensor Type And Correction Applied To Current Measurement - Agilent Technologies E4419B User Manual

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

The softkey menu structured to this key affects the channel setup.
Press this key to access the "Frequency/Cal Fac" menu which allows you to
enter the frequency of the signal to be measured or the calibration factor.
Depending on the type of power sensor connected and whether or not a sensor
calibration table or frequency dependent offset table is selected, the text on
some softkeys may appear grayed out. This indicates that they are not relevant
in the power meter's current mode of operation.
softkeys in various modes of operation.

Table 3-1 Combination of Sensor Type and Correction Applied to Current Measurement

Sensor Model
Sensor Calibration Table Selected
Frequency
A Freq
B Freq
Agilent 8480
Frequency can be
Series power
entered. Default
sensor
value is 50 MHz.
Agilent E-Series
Not Applicable
power sensors
Agilent N8480
Not Applicable
Series power
sensors (excluding
Option CFT)
Agilent N8480
Frequency can be
Series power
entered. Default
sensors with
value is 50 MHz.
Option CFT
E4419B Power Meter User's Guide
Cal Factor
A Cal Fac
B Cal Fac
The Cal Factor is
obtained from the
sensor calibration table
but can be overridden
by this softkey.
Not Applicable
Not Applicable
The Cal Factor is
obtained from the
sensor calibration table
but can be overridden
by this softkey.
Menu Reference
Table 3-1
details the relevant
Sensor Calibration Table Not Selected
Frequency
Cal Factor
A Freq
A Cal Fac
B Freq
B Cal Fac
Frequency can be
The Cal Factor can be
entered if a
entered. Default value
frequency
is 100%.
dependent offset
table is selected.
Frequency can be
The Cal Factor cannot
entered. Default
be entered, it is
value is 50 MHz.
obtained from the
EEPROM in the power
sensor.
Frequency can be
The Cal Factor cannot
entered. Default
be entered, it is
value is 50 MHz.
obtained from the
EEPROM in the power
sensor.
Frequency can be
The Cal Factor can be
entered if a
entered. Default value
frequency
is 100%.
dependent offset
table is selected.
3
127