2
Power Meter Operation
Setting Offsets
Setting Channel Offsets
The power meter can be configured to compensate for a signal loss or gain in
your test setup (for example, to compensate for the loss of a 10 dB attenuator).
This gain or loss referred to as a channel offset, is added to the measured
power before the result is displayed.
Offsets are entered in dB. The allowable range of values is –100 dB to +100 dB.
A positive value compensates for a loss and a negative value compensates for a
gain.
If either a channel or a display offset is set, then "Ofs" is displayed.
To enter a channel offset which compensates for a loss or gain:
1 Press
2 Press
3 Press
4 To confirm your choice, press
5
Channel offsets can be disabled and re-enabled simply by pressing
Offset Off On
Setting Display Offsets
The display offset function provides a method for entering display offset
values.
66
.
A Input Settings
which you want to apply offsets to.
. The power meter displays the offset value in a pop-up window.
Offset
Modify this offset (see below) as desired.
• Use
or
to modify the digit on which the cursor is currently
positioned.
• Use
or
to move to other digits.
Offset Off On
is automatically set to "On" when a value is entered using
.
Offset
.
or
B Input Settings
.
dB
depending on the channel
E4419B Power Meter User's Guide