Table 5-6 Settling Time - Agilent Technologies E4419B User Manual

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

5
Specifications
For Agilent N8480 Series power sensors
Manual filter, 10 dB decreasing power step:

Table 5-6 Settling Time

Number of Averages
Settling Time (s)
(Normal Mode)
Response Time (s)
(x2 mode)
Auto filter, default resolution, 10 dB decreasing power step,
normal and x2 speed modes:
Normal Mode
Maximum dBm
150 ms
Typical
200 ms
Settling
Times
1 s
6.6 s
13.5 s
Minimum dBm
176
1
2
4
8
0.15
0.2
0.3
0.5
0.15
0.18
0.22
0.35
20 dB
Power
10 dB
Sensor
Dynamic
Range
10 dB
10 dB
5 dB
16
32
64
128
1.1
1.9
3.4
6.6
0.55
1.1
1.9
3.5
x2 Mode
Maximum dBm
150 ms
Typical
180 ms
Settling
Times
400 ms
3.6 s
6.6 s
Minimum dBm
E4419B Power Meter User's Guide
256
512
1024
13
27
57
6.9
14.5
33
20 dB
Power
10 dB
Sensor
Dynamic
Range
10 dB
10 dB
5 dB

Hide quick links: