Presetting The Power Meter; Preset Conditions - Agilent Technologies E4419B User Manual

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

Presetting the Power Meter

This section details the power meter's preset conditions.
The GPIB address, the data stored in the sensor calibration tables and the
zeroing and calibration data are not affected by a preset. The calibration table
selected is not affected.

Preset Conditions

The number of windows displayed is two.
Measurement units used is set to dBm.
E4419B Power Meter User's Guide
is set to 50 MHz.
A Freq
is set to 100%.
A Cal Fac
is set to 50 MHz.
B Freq
is set to 100%.
B Cal Fac
is set to Channel A in the upper window and Channel B in the
Input Select
lower window.
is set to digital for both windows.
Display Select
is set to 20.000 dBm (
Anlg Mtr Scaling
Resolution 1 2 3 4
is set to "Off"
Limits Off/On
is set to 90.000 dBm (
Limits
TTL Output Off/On
Limits OVER, UNDER, EITHER
Fail O/P High/Low
is set to "3".
) and –90.000 dBm (
Max
is set to "Off".
is set to "OVER".
is set to "Low"
Power Meter Operation
) and –70.000 dBm (
Max
).
Min
2
).
Min
97