E4419B Power Meter User's Guide
Insert
Press this softkey to insert a new table entry. You are prompted to
enter the frequency and calibration factor or offset. The entry is
inserted in ascending frequency order.
Delete
Press this softkey to delete a table entry. If you delete the
frequency, the corresponding calibration factor or offset is also
removed and vice versa.
Done
Press this softkey to return to the measurement screen. The
softkeys from the first level of the "System/Inputs" menu are
displayed.
A Table Off On
Press this softkey to enable and disable the highlighted table for
channel A. "A" or "B" is displayed beside each table to indicate its
current state and use.
B Table Off On
Press this softkey to enable and disable the highlighted table for
channel B. "A" or "B" is displayed beside each table to indicate its
current state and use.
Done
Press this softkey to return to the measurement screen. The softkeys
from the first level of the "System/Inputs" menu are displayed.
A Linearity AType DType
Press these softkeys to select the type of linearity correction that will be
applied to the channel A and channel B sensors being used. For most
8480 series sensors, the correct (A type or D type) linearity correction
table is automatically selected. However, for the V8486A and W8486A
sensors, the automatic selection must be overridden and the D type
correction selected. Subsequent connection of another A type sensor will
result in a warning message stating that "Linearity Override May be
Required".
A Input Settings
and
Press these softkeys to access a menu which allows you to modify the
settings of the averaging, duty cycle, range and offset of channel and
channel B respectively.
and
B Linearity AType DType
B Input Settings
Menu Reference