Thermocouple Measurement/Simulation, Connections and
Troubleshooting

Internal Reference Junction

MC5-IS' Internal Reference Junction Mod-
ule is an optional module. To be able to
utilize Internal Reference Junction com-
pensation this optional module has to be
installed into your MC5-IS. If the measur-
ing/simulating port is set to the Internal
Reference
Junction
ET: TCi(mea)
(
or
automatically selects the Internal Refer-
ence Junction compensation method. No
other Reference Junction compensation
methods are available unless the mea-
suring/simulating port is changed.
See also...
To accurately measure the thermovoltage caused by the tempera-
ture to be measured, the second thermovoltage caused by the
Reference Junction needs to be compensated. This is done us-
ing one of the Reference Junction compensation methods de-
scribed in the subsequent chapters.
The Reference Junction compensation method has to be chosen
both when measuring and simulating thermocouples.
Module
ET: TCi(sim)
), MC5-IS
Note.
The range of the Internal Reference Junction's temperature com-
pensation is -10 ... +50°C (14 ... 122 °F).
External Reference Junction on page 90
Additional Information
Connection when measuring/simulating
thermovoltage using Internal Reference
Junction Mode:
T/C materials
(T/C, extension or
compensation wires)
T/C sensor
or a
T/C signal receiver
T/C INT. RJ
4 w meas
T/C, Low V
R, RTD
T/C
T/C
WIRES
WIRES
ONLY
ONLY
meas/sim
3 w meas
SENSOR MEASURE & SIMULATE
89