HART Instruments and Beamex's Calibration Database Software

Connecting MC5-IS and a HART
Instrument

External Supply

130
Appendix 1, User Guide for MC5 HART Option
Beamex's QCAL
®
Calibration Database Software support HART in-
struments as follows:
QM6 Quality Manager starting from version 1.90.
adding instruments created in MC5-IS to QM6's database,
starting from version 1.90a.
QD3 Quality Documenter starting from version 2.10.
How the HART instrument is connected to MC5-IS depends on the
supply power connection and whether there is a test diode con-
nected to the transmitter. The following pictures describe the con-
nections for each case.
MC5-IS measures the externally supplied output current of the trans-
mitter. The HART
terminal is used for
digital communica-
tion. Remember to
make sure that the
loop also includes a
resistor with a resis-
tance between 230 to
600 ohm or that the
impedance of the
loop itself is at least
230 ohms.
®
X
Input signal
T/C
INT. RJ
T/C, Low V
T/C
WIRES
ONLY
SENSOR MEASURE & SIMULATE
+
I
4 w meas
HART
HART
I
I
®
®
meas/sink
meas/sink
R, RTD
V,
I
,
sink
V,
,
Com
Com
meas/sim
3 w meas
Uo: 16 V
Io: 24 mA
Po: 145 mW
OUTPUT
MEASURE
ET
E
Low V
Supply
etc.