Temperature Indicators and Recorders

Required modules
Preparations
1. Connect suitable terminals in MC5-IS
to the input connectors of the indicator/
recorder.
2. If the sensor to be simulated is a ther-
mocouple, select one of the following
reference junction methods:
The internal Reference Junction
module is in use.
The reference junction is fixed to
0°C.
The reference junction tempera-
ture is entered.
The reference junction tempera-
ture is measured externally using
an RTD connected to the RTD ter-
minals.
See Part C for additional information
concerning the reference junction meth-
ods.
3. If the sensor to be simulated is an RTD,
the used wiring system depends on the
instrument to be calibrated. Remember
to only use the two leftmost RTD termi-
nals in MC5-IS.
114
Calibration
This procedure suits Temperature Indicators/Recorders no matter
if the sensor is an RTD or a thermocouple. The instruments input
signal is simulated using MC5-IS and the output reading is entered
into MC5-IS.
The ET module for simulating the temperature sensor. The
Base Unit takes care of entered output signal.
4. Test the connections in Basic Mode if
needed. To quickly configure the Basic
Mode's windows, go to Calibration
Mode, select the instrument to be cali-
brated and the Function Key
brate
but immediately return to Basic
Mode.
T/C INT. RJ
T/C, Low V
R, RTD
T/C
WIRES
ONLY
meas/sim
SENSOR MEASURE & SIMULATE
?
06.10.2000 10:46
Input
RTD Temperature [ET: Simul.]
75.00
Output
Value [Keyed]
74.800
Enter Output Reading
Error
1.00
II 1 G
+
EEx ia IIC T4
(Tamb = 50°C)
VTT 02 ATEX 003 X
0
Ex ia IIC T4
(Tamb = 50°C)
VTT No. Ex-02.011 X
-
0%
INPUT
Accept
Ui: 30 V
Pause
Readings
Ii: 215 mA
Pi: 1 W
M U L T I F U N C T I O N
Cali-
C
/
HART
I
4 w meas
®
meas/sink
I
V,
,
sink
V,
,
Com
3 w meas
Uo: 16 V
Io: 24 mA
Po: 145 mW
OUTPUT
MEASURE
ET
E
Low V
Pt100 α385
°C (ITS90)
°C
-0.20 % of span
50%
100%
C A L I B R A T O R