Device Settings

Trimming a HART Instrument in Basic Mode

HART and MC5-IS' Basic Mode
The adjacent picture displays the
data of the Device Settings window.
The following fields are editable:
Tag,
Polling Address,
Lower Range,
Upper Range,
Damping,
Transfer Function,
Burst Mode
Sensor Serial Number.
Note that MC5's Calibration Mode
also includes a possibility to trim
HART instruments. Trimming a
HART Instrument in Basic Mode is
only needed when you do not in-
tend to calibrate and trim the
HART instrument.
To start trimming the HART
instrument, press:
Menu
D
/
and
Window 2 Setup
, if needed
Trim Instrument
6
/
From the pop-up list, select which
output, analog or digital, is to be
trimmed. MC5-IS reads the
instrument data while displaying
the following message:
Reading HART instrument data.
Please wait.
and
.
11.04.2002 8:38
DEVICE SETTINGS
Tag
3051C
Polling Address
0
Output
Range Unit
mbar
Lower Range
0.00
Upper Range
1000.00
Damping (s)
0.40
Transfer Function
Linear
Burst Mode
Off
Sensor
Serial Number
1109949
Lower Limit
0.00
Upper Limit
2486.41
Minimum Span
24.87
Close
11.04.2002 8:49
1
Pressure
Disconnect
P1: INT400MC/-41.400...41.400 kPa
Device
gauge
0.088
Device
kPa
Information
Device
Settings
Unit
[kPa]
2
Pressure
E: HART/PV
0.084
kPa
Analog Output (AO)
Trim
Digital Output (PV)
Instrument
Add HART
Instrument
to Database
Window 1
Window 2
Others
Setup
Setup
Close
Close
MENU
135