Calibration Result Windows; How To Choose Which Calibration Run Is Viewed; Deleting Calibration Results - BEAMEX MC5-IS User Manual

Nstrinsically safe multifunction calibrator
Hide thumbs Also See for MC5-IS:
Table of Contents

Advertisement

Calibration Result Windows

How to Choose Which Calibration Run is Viewed

Deleting Calibration Results

Viewing Calibration Results
The results are displayed in table
format and as a graph. There are
also some additional data display-
ing the statistics of the calibration
as well as environmental informa-
tion that was automatically inserted
during the calibration or it was
manually entered after the calibra-
tion. If any calibration notes were
written, they can also be seen on
one of the calibration result pages.
Next Page
B
Use the
/
from one result page to another.
The table with numeric results may
include more rows in the table than
what can be displayed. Use the
vertical cursor keys to scroll
through all rows.
When viewing Calibration results,
MENU
D
press
/
,
History
. A list of saved calibrations
is displayed. Select the calibration
run to be viewed by moving the
cursor (inverted text) to the desired
calibration run and pressing either
the
or the
key. .
When viewing Calibration results, press
To remove the current calibration run, press
To remove all calibration results (for the current instrument), press
Remove All Results
4
/
, to change
Calibration
1
/
.
03.10.2000 15:22
Output Error [% of span]
0.40
+
0
-
0%
50%
Maximum Error
Maximum Hysteresis
Maximum Unlinearity
Span Error
Zero Error
Next
Back
Page
03.10.2000 15:22
25.09.2000 10:52 - As Left - Passed
Input
Output
[°C]
[mA]
0.00
3.9930
25.00
7.9890
50.00
11.9834
75.00
15.9777
100.00
19.9729
75.00
15.9875
50.00
11.9925
25.00
7.9979
0.00
4.0032
1
Next
Back
Page
MENU
D
/
.
Remove Results
3
/
100%
-0.170
0.064
0.044
-0.126
-0.044
MENU
Error
[%]
-0.044
-0.069
-0.104
-0.139
-0.170
-0.078
-0.047
-0.013
0.020
2
MENU
.
125

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents