Viewing Calibration Results

124
Calibration
You can open the calibration result windows from two places:
When viewing the instrument window, press
Calibration Results
03.10.2000 15:21
INSTRUMENT
112-TT-003.1
Intake Temperature on Feed 1
Transfer Function Linear
5 ↑↓
Cal. Points
INPUT
0.00 ... 100.00 °C
Method
Simulated
Pt100 α385
Sensor Type
Output
4.0000 ... 20.0000 mA
Method
Measured
Back
Edit
Calibrate
When viewing the calibration windows, and you haven't started a
calibration run, press
(see the rightmost picture above).
(see the leftmost picture below).
View
Input
Instrument
Details
Delete
Instrument
Output
View
Calibration
Results
Error
1.00
+
0
-
Close
MENU
MENU
Menu
View Calibration Results
D
4
/
and
/
Menu
D
/
and
03.10.2000 15:22
Zero
RTD Temperature [ET: Simul]
Pt100 α 385
Pressure
0.00
Module
°C (ITS90)
Prescan
[On]
Current [E: Meas]
3.9762
mA
View
Calibration
Results
-0.15 % of span
0%
50%
100%
Close
Back
Start
Adjust
MENU
MENU
View
3
/