Starting MC5-IS

Startup Procedure

32
Startup and Basic Operation
When MC5-IS is started, a startup picture appears. After a self test,
some basic information of the calibrator at hand appears in the lower
part of the screen. If you want to view the calibrator information for
a longer period, press the
information is visible until you press the
S e r i a l n u m b e r
M a i n v e r s i o n
E m o d u l e v e r s i o n
E T m o d u l e v e r s i o n
C a l i b r a t i o n d u e d a t e
If a module's version number cannot be seen, the module is not
included in the MC5-IS at hand.
The calibration due date that is listed in the startup window is the
earliest calibration due date for all connected modules. If the cali-
brator requires recalibration, MC5-IS stops at the calibrator infor-
mation window and the text "
calibration due date row.
Wait
/
function key. Then the calibrator
D
D
w w w . b e a m e x . c o m
Calibrate Soon
Continue
/
function key.
" appears below the