A Calibration Procedure Using MC5-IS

106
Calibration
The Calibration Windows:
Input
Signal
Window
Output
Signal
Window
Error
Graph
A typical calibration procedure using MC5-IS is as follows:
We assume you have already selected the instrument to be
calibrated and are viewing/editing the selected instrument's data
pages. See Selecting the Instrument to Be Calibrated on page
104 for information on how to enter Calibration Mode and se-
lecting instruments for calibration.
1. Make the required connections and test them by altering the
input signal in with the Instrument Adjustment utility (if you
haven't done the connections and testing in Basic Mode).
2. Start the calibration run by pressing
IS passes through all calibration points depends on the
Calibration Method
In Automatic calibration MC5-IS generates or simulates
the input signal. See chapters
About Automatic Calibration on page 108 and
About Manual Calibration on page 108.
Note.
You can pause a calibration run whenever you want by
A
pressing the
with the following (self-explanatory) choices:
Continue
A
/
End
B
/
Skip Point
C
/
25.09.2000 10:46
Input
RTD Temperature [ET: Simul]
0.00
Output
Current [E: Meas]
3.9762
Error
1.00
+
0
-
0%
Back
Start
setting:
Stop
/
Function Key. Then you are provided
.
Pt100 α385
°C (ITS90)
mA
-0.15 % of span
50%
100%
Adjust
MENU
Start
B
/
. The way MC5-