Electrical Module (E)

Function
(2
mV measurement
(3
V measurement
(4
mA measurement
(5
Hz measurement
(5
Pulse counting
mA sink
Uncertainty includes reference standard uncertainty, hysteresis, nonlinearity, repeatability and typical long term stability for mentioned
1 )
period. (k=2)
2 )
Bias current <10 nA
3 )
Impedance >1 Mohm
4 )
Impedance < 7.5 ohm
5 )
Impedance > 1 Mohm. Frequency measurement minimum amplitude 1 Vpp (< 10 kHz),
3 Vpp (10...50 kHz). Pulse counting minimum amplitude 1 Vpp (pulse length > 50 µs),
3 Vpp (pulse length 50 µs...10 µs).
Trigger level range -1...+15 V.
Range
±250 mV
±30 V
±100 mA
0.0028 to 50000 Hz
0 to 9 999 999 pulses
0 to 25 mA
Modules
Resolution
0.001 mV
0.00001 - 0.001 V
0.0001 - 0.001 mA
0.000001 - 0.1 Hz
1 pulse
0.0001 mA
All specifications are subject to change without prior notice
(1
1 Year Uncertainty
0.02 % RDG + 5 µV
0.02 % RDG + 0.25 mV
0.02 % RDG + 1.5 µA
0.01 % RDG
N/A
0.02 % RDG + 1.5 µA