Calibrating A Hart Instrument; Selecting The Instrument To Be Calibrated - BEAMEX MC5-IS User Manual

Nstrinsically safe multifunction calibrator
Hide thumbs Also See for MC5-IS:
Table of Contents

Advertisement

Calibrating a HART Instrument

Selecting the Instrument to be Calibrated

138
Appendix 1, User Guide for MC5 HART Option
To be able to calibrate the HART
instrument, it has to be added to
MC5-IS' instrument database. The
analog output (AO) and the digital
output (PV) are treated as separate
instruments.
Adding the instruments can be
done using the utility described in
chapter Adding a HART Instru-
ment to MC5-IS' Instrument Da-
tabase on page 137 or by sending
the instrument data from a calibra-
tion software. Additionally: you can
also manually enter the instrument
data into MC5-IS' instrument data-
base.
Select the instrument to be calibrated from MC5-IS' list of available
instruments.
If you are already connected to the
HART instrument (HART commu-
nication started in Basic Mode),
MC5-IS continues directly from the
Instrument Window to the Calibra-
tion Windows.
If HART communication is not
started and you are calibrating the
digital output (PV) of the HART in-
strument, MC5-IS prompts you to
start the communication and select
the instrument in a window similar
to the adjacent picture.
MC5-IS accepts the HART instru-
ment as the instrument to be cali-
brated only if its instrument data matches with the data of the instru-
ment selected in MC5-IS.
06.10.2000 17:13
INSTRUMENT
112-TT-003.1
JUST TESTING
Transfer Function Linear
5 ↑↓
Cal. Points
INPUT
0.00 ... 1000.00 mbar
Method
Measured
Pressure Type
Gauge
OUTPUT
4.0000 ... 20.0000 mA
Method
Measured
Back
Edit
Calibrate
06.10.2000 17:16
HART
Searching for HART Device at:
Devices found:
0 - 3051 - Pressure Transmitter
1 - PT155 NC-TTR
2 - PT311 ARG - JT
3 - PT413 WBA4 - NTB
4 - TT210 PLGR - NSWF
5 - TT921 FPN - ILL
6 - TT148 TBB - NE
Back
Restart
MENU
8