HART and MC5-IS' Basic Mode

Connecting to a HART Instrument

HART and MC5-IS' Basic Mode
To communicate with a HART in-
strument, press:
Menu
D
/
and
Window 2 Setup
B
/
HART
5
/
.
Note.
The HART menu option is disabled if
the other window already reserves the
E module. Change the setup of the other
window to enable the HART menu op-
tion.
Refer to chapter Connecting MC5-
IS and a HART Instrument on
page 130 for information on how to
connect MC5-IS and a HART in-
strument.
MC5-IS searches for HART devices
connected to the current loop and
opens a window similar to the one
seen to the right.
Up to 16 instruments may be con-
nected to the same current loop.
Choose the appropriate instrument
Select
4
by pressing
/
ther the
or the
Note.
Measuring the Analog Output (AO) of a
HART instrument does not require any
special actions. Just connect the Analog Output to MC5-IS' current mea-
surement terminals in the E module.
, if needed
(optionally ei-
key).
11.04.2002 8:25
1
Pressure
Quantity
P1: INT400MC/-41.400...41.400 kPa
[Current]
gauge
0.089
Function/Port
kPa
[E:I(Meas)]
Display Mode
[Eng. Unit]
Unit
[mA]
2
Current
E: Current Measurement
HART
3.9855
mA
Window 1
Window 2
Others
Setup
Setup
11.04.2002 8:26
HART
Searching for HART Device at:
Devices found:
0 - AA377039 -Pressure Transmitter
1 - PT155 NC-TTR
2 - PT311 ARG - JT
3 - PT413 WBA4 - NTB
4 - TT210 PLGR - NSWF
5 - TT921 FPN - ILL
6 - TT148 TBB - NE
Back
Restart
Close
MENU
12
Select
133