Thermocouple Simulation

MC5-IS is able to mimic a thermocouple at given temperatures. This property is called
thermocouple simulation. It enables checking and calibration of temperature indicators,
recorders, transmitters and other equipment related to temperature measurement with a
thermocouple probe.
In thermocouple simulation the original thermocouple is disconnected from the instru-
ment under test and replaced with MC5-IS. To the instrument under test MC5-IS appears
as a thermocouple at the given temperature.
Thermocouple simulation terminals are located in the
Required settings
Quantity
Funct/Port
Sensor Type
Reference Junction

Internal Reference Junction

MC5-IS' Internal Reference Junction Mod-
ule is an optional module. To use the Inter-
nal Reference Junction, select Function/
ET: TCi(sim)
Port
. The Reference Junction
compensation method
matically selected.
T / C
I N T . R J
T / C , L o w
V
R , R T D
T / C
T / C
W
W
I R E S
I R E S
O N L Y
O N L Y
m e a s / s i m
S E N S O R
M E A S U R E
&
S I M U L A T E
Additional information is found in chapter
Internal Reference Junction on page 91.
Options/description
T/C-Temp
ET: TCi(sim)
(for the internal Reference Junction), or
ET: TCx(sim)
(for other RJ compensation methods)
Available thermocouple types
Depends on the Function/Port setting.
See subsequent chapters.
Internal
is auto-
1
4
w
m e a s
H A R T
®
1


m e a s / s i n k
V ,
 
 
,
s i n k
V ,
,
C o m
3
w
m e a s
U o : 1 6
V
I o : 2 4
m A
P o : 1 4 5
m W
O U T P U T
M E A S U R E
E T
E
L o w
V
Generating/Simulating
module.
ET

External Reference Junction

To use an External Reference Junction,
select Function/Port
choose one of the available Reference
Junction compensation methods:
0°C
or available
T / C
I N T . R J
T / C , L o w
T / C , L o w
V
V
T / C
W
I R E S
O N L Y
S E N S O R
M E A S U R E
Additional information is found in chapter
External Reference Junction on page 92.
ET: TCx(sim)
Entered
RTD sensors
.
1
4
w
m e a s
H A R T
®
R , R T D
1


m e a s / s i n k
V ,
 
 
,
s i n k
V ,
,
C o m
m e a s / s i m
3
w
m e a s
U o : 1 6
V
I o : 2 4
m A
P o : 1 4 5
m W
O U T P U T
M E A S U R E
&
S I M U L A T E
E T
E
L o w
V