Electrical Limit Switches - BEAMEX MC5-IS User Manual

Nstrinsically safe multifunction calibrator
Hide thumbs Also See for MC5-IS:
Table of Contents

Advertisement

Electrical Limit Switches

Required modules
Preparations
1. Connect the ET module's terminals
OUTPUT
marked "
switch. Add a current source to the loop.
MC5-IS controls the current provided by
the current source.
2. Connect the switch contact to the switch
terminals in the E module.
3. Test the connections in Basic Mode if
needed. To quickly configure the Basic
Mode's windows, go to Calibration
Mode, select the instrument to be cali-
brated and the Function Key
brate
but immediately return to Basic
Mode.
Current
Source
T/C INT. RJ
T/C, Low V
T/C
WIRES
ONLY
SENSOR MEASURE & SIMULATE
Input
Testing switch limits (4/5)
Output
II 1 G
EEx ia IIC T4
(Tamb = 50°C)
VTT 02 ATEX 003 X
Ex ia IIC T4
(Tamb = 50°C)
VTT No. Ex-02.011 X
INPUT
Ui: 30 V
Pause
Ii: 215 mA
P i: 1 W
M
U
L
T
I
F
U
116
Calibration
This procedure suits limit switches with an electrical input. The in-
put signal is generated with MC5-IS.
The ET module for generating the required electrical signal
(Voltage or Current).
The E module for detecting the switch's state.
" to the input of the
4 w meas
HART
®
I
R, RTD
I
meas/sink
V,
,
sink
V,
,
Com
meas/sim
3 w meas
Uo: 16 V
Io: 24 mA
Po: 145 mW
OUTPUT
MEASURE
ET
E
Low V
?
06.10.2000 10:29
Voltage [ET: Gen.]
5.023
V
Switch [E: Meas.]
8.0310
Actuate
-------
Deactuate
7.000
5.500
4.500
3.000
N
C
T
I
O
N
C A L I B R A T O R
Calibration
1. Move to Calibration Mode and select the
instrument to be calibrated.
2. Start the calibration as is presented in
chapter A Calibration Procedure Us-
ing MC5-IS on page 106. MC5-IS does
Prescan
the
prescan, MC5-IS searches for approxi-
mate values for the actuating and
deactuating point. This speeds up the
final test without sacrificing accuracy.
The prescan test is done only once in a
calibration.
Cali-
C
/
By default, the
If you do not want MC5-IS to per-
form a prescan, disable it (com-
mands
When
IS uses the whole scan range also
during the actual test.
3. The actual test is done automatically:
MC5-IS slowly increases the input sig-
nal until the switch actuates and con-
tinues by decreasing the input signal
until the switch deactuates. MC5-IS'
screen displays the obtained data as
the test advances.
if it is enabled. During the
Prescan
is set to
MENU
Prescan
D
2
/
and
/
Prescan
off
is set to
on
.
).
, MC5-