Phases Of Instrument Calibration - BEAMEX MC5-IS User Manual

Nstrinsically safe multifunction calibrator
Hide thumbs Also See for MC5-IS:
Table of Contents

Advertisement

Phases of Instrument Calibration

Start
Read Starting
Guide Text
As Found test
Save As
Found results
No. of
yes
no
repeats
done?
Adjustment
no
required?
* If it is not possible
to adjust the
maximum error
below the "Reject if"
limit, consider
replacing the
instrument with
a more accurate
one.
The picture gives a general view of the phases in a calibration
procedure. MC5-IS supports three separate guide texts (white
blocks in the picture):
One is displayed when starting the calibration.
Another when the adjustment utility is started.
The third text is displayed when ending the calibration.
The guide texts may be entered in MC5-IS or they may be down-
loaded from a PC with a calibration software.
General
yes
Read Adjusting
Guide Text
Adjustment
Within
yes
no*
limits?
As Left test
Save As Left
results
No. of
no
repeats
done?
yes
Read Finishing
Guide Text
End
99

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents