Phases of Instrument Calibration
Start
Read Starting
Guide Text
As Found test
Save As
Found results
No. of
yes
no
repeats
done?
Adjustment
no
required?
* If it is not possible
to adjust the
maximum error
below the "Reject if"
limit, consider
replacing the
instrument with
a more accurate
one.
The picture gives a general view of the phases in a calibration
procedure. MC5-IS supports three separate guide texts (white
blocks in the picture):
One is displayed when starting the calibration.
Another when the adjustment utility is started.
The third text is displayed when ending the calibration.
The guide texts may be entered in MC5-IS or they may be down-
loaded from a PC with a calibration software.