Calibration Settings Page - BEAMEX MC5-IS User Manual

Nstrinsically safe multifunction calibrator
Hide thumbs Also See for MC5-IS:
Table of Contents

Advertisement

Calibration Settings Page

122
Calibration
Switches have an altogether differ-
ent group of settings as shown in
the adjacent picture.
The actuation and deactuation
point errors are calculated against
the corresponding nominal values.
Error Calculation Directions
fine error directions to be observed
when calculating errors.
Scan Range
values define the
range where MC5-IS searches for
the approximate actuation and
deactuation point at the beginning
of an automatic calibration of a
switch. The range limits should be symmetrical in relation to the
actuation/deactuation points.
Setpoint Delay
The
use only when the
Method
is set to
defines how long MC5-IS waits be-
fore saving the input and output
values after input value was
changed to the next calibration
point.
Maximum Allowed Calibration
Point Deviation
is a limit that de-
fines how far the input signal value
of the saved point can be from the
theoretical calibration point. This
setting is used both in manual and
automatic calibration.
Again, the settings for a switch is somewhat different:
Number of Repeats
An additional
tion Points
Maximum Allowed Calibration Point Deviation
or
fields.
de-
setting is of
Calibration
Automatic
. It
field is visible but no
03.10.2000 13:05
INSTRUMENT OUTPUT
Switch: Pressure [bar]
Switch Type
Normally Closed
Nominal Points
Actuating
0.700000
Deactuating
0.600000
Error Calculation Directions
Actuating
Up/Down
Deactuating
Up/Down
Scan Range
Start Point
0.400000
End Point
0.900000
Next
Save
Edit
page
04.01.2002 12:52
CALIBRATION SETTINGS
Calibration Method
Manual
Calibration Points
5
Setpoint Delay
5.00
Maximum Allowed Calibration Point
Deviation
<
0.00
Calibration Repeats
As Found
0
As Left
1
Cal. Period, Days
<
360
Next
Save
Edit
page