Scaling - BEAMEX MC5-IS User Manual

Nstrinsically safe multifunction calibrator
Hide thumbs Also See for MC5-IS:
Table of Contents

Advertisement

Scaling

78
Advanced Operation and Configuration
In scaling, the measuremed/generated/simulated value is dis-
played in another, user defined quantity/unit. Start scaling as de-
scribed in chapter Display Mode Settings on page 77.
When selecting Scaling Display Mode, the following configuration
window appears:
26.04.2002 14:02
SCALING
Quantity
Pressure
Port
P1:INT2C
Measured Range
0 %
0.000
100 %
100.000
Current
Scaled Mode
Scaled Unit
mA
Scaled Range
0 %
4.000
20.000
100 %
Transfer Function
Linear
Stop
Cancel
Field
Ramping
You can scale the quantity either to a quantity that already exists in
MC5-IS or select
Scale Mode
from the
you can manually enter any unit to
Scale Unit
the
leftmost picture above).
When scaling is active in a Basic
Mode window, the window con-
SCALED
tains the text "
value is displayed in Extra Info row.
If the Extra Info row was assigned
to some other measurement be-
fore scaling was started, the other
measurement is automatically
stopped.
Note.
Remember to set the amount of decimals to be displayed for a
Custom Unit. The default value is 3.
kPa
kPa
Edit
Start
Custom Units
field. Then
field (see the
". The true
30.05.2002 10:09
SCALING
Quantity
Pressure
Port
P1:INT2C
Measured Range
0 %
0.000
100 %
100.000
Custom Units
Scaled Mode
Scaled Unit
Gallons/min
Scaled Range
0 %
1.00
5.00
100 %
3
Decimals
Transfer Function
Linear
Stop
Cancel
Field
Edit
Ramping
30.05.2002 10:10
1
Pressure
P2: INT20C/-1.0 ... 20.7 bar g
SCALED
20.088
Gallons/min
Measurement bar
2
Current
E: Current Measurement
8.3217
mA
Calibration
Mode
kPa
kPa
Start
1.004
MENU

Hide quick links: