Adding a HART Instrument to MC5-IS' Instrument Database

HART and MC5-IS' Basic Mode
MC5-IS has a semi-automatic util-
ity to add HART instruments to
MC5-IS' instrument database.
Add HART instrument
7
Select
/
to Database
from the Basic
Mode's HART Menu. Continue by
selecting which output is to be
added, analog or digital.
The following instrument data is
acquired from the HART instru-
ment:
Input Quantity
Input Range
Output Quantity
Output Range
Transfer Function Device ID
Setpoint Delay
When the instrument is created,
MC5-IS automatically opens the
Instrument Window, from where it
is easy to either check/edit instru-
ment data or start calibrating the
instrument.
It is recommended to edit the instru-
ment data and carefully check the
default values assigned for instru-
ment data fields that do not have
corresponding fields in HART in-
strument data. To edit the instru-
ment data, press the
At least the following instrument data fields should be checked:
Error Calculation Method
Calibration Method
Keep in mind that changing the value of some of the above men-
tioned fields may also require changes in related fields.
Notes.
The HART instrument's TAG is assigned to the Device ID field in MC5-IS.
No Position ID is automatically assigned to the instrument.
QM6 Quality Manager version 1.90a and later support adding instruments
created in MC5-IS to QM6's database. QD3 Quality Documenter cannot
receive instrument data from MC5-IS. It can receive only calibration results
of instruments already existing in QD3's database.
Input Unit
Input Method
Output Unit
Output Method
Edit
2
/
button.
Reject if >
Calibration Points
11.04.2002 9:57
1
Pressure
Disconnect
P1: INT400MC/-41.400...41.400 kPa
Device
gauge
1.1
Device
kPa
Information
Device
Settings
Unit
[mbar]
2
Pressure
E: HART/PV
1.142
kPa
Trim
Instrument
Add HART
Analog Output (AO)
Instrument
Digital Output (PV)
to Database
Window 1
Window 2
Others
Setup
Setup
06.10.2000 16:57
INSTRUMENT
112-TT-003.1
JUST TESTING
Transfer Function Linear
3 ↑
Cal. Points
INPUT
0.00 ... 1000.00 mbar
Method
Measured
Pressure Type
Gauge
OUTPUT
4.0000 ... 20.0000 mA
Method
Measured
Back
Edit
Calibrate
Input Port
Output Port
Close
MENU
MENU
137

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents