Current Measurement - BEAMEX MC5-IS User Manual

Nstrinsically safe multifunction calibrator
Hide thumbs Also See for MC5-IS:
Table of Contents

Advertisement

Current Measurement

The current measurement terminals are located in the
100 mA.
Required settings
Quantity
Funct/Port
Next...
Options/description
Current
E: I(meas)
The active terminals are shown in the picture below.
T / C
I N T . R J
T / C
W
I R E S
O N L Y
Notes.
Check the polarity of your connections. The arrows in the previous
picture describe the correct flow of current.
If you are using a transmitter with HART capability, make also a
connection to MC5-IS' HART terminal. For more information con-
cerning HART connections, see Appendix 1, chapter Connecting
MC5-IS and a HART Instrument on page 131. Information con-
cerning current measurement parallel to a test diode can be found
in part C, chapter Current Measurement Parallel to a Test Diode,
Connections on page 97.
Current Sink on page 58
Special Measurements on page 51.
Alarm Limit Settings on page 71.
Calibration, see Part D.