Current Measurement
The current measurement terminals are located in the
100 mA.
Required settings
Quantity
Funct/Port
Next...
Options/description
Current
E: I(meas)
The active terminals are shown in the picture below.
T / C
I N T . R J
T / C
W
I R E S
O N L Y
Notes.
Check the polarity of your connections. The arrows in the previous
picture describe the correct flow of current.
If you are using a transmitter with HART capability, make also a
connection to MC5-IS' HART terminal. For more information con-
cerning HART connections, see Appendix 1, chapter Connecting
MC5-IS and a HART Instrument on page 131. Information con-
cerning current measurement parallel to a test diode can be found
in part C, chapter Current Measurement Parallel to a Test Diode,
Connections on page 97.
Current Sink on page 58
Special Measurements on page 51.
Alarm Limit Settings on page 71.
Calibration, see Part D.
Measuring
module. The maximum current is
E
4
w
m e a s
H A R T
H A R T
T / C , L o w
V
R , R T D
1
V ,
,
s i n k
m e a s / s i m
V ,
,
3
w
m e a s
U o : 1 6
V
I o : 2 4
m A
P o : 1 4 5
m W
O U T P U T
M E A S U R E
S E N S O R
M E A S U R E
&
S I M U L A T E
E T
E
1
1
®
®
m e a s / s i n k
m e a s / s i n k
C o m
C o m
L o w
V
39