Special Filtering and Resolution

Deviation Measurement

Start using special filtering and
resolution by selecting:
Menu
/
D
Window 2 Setup
and
/
B
needed,
or
,
Special Measurements
/
1
Filter and Resolution...
Select the Time Constant setting
and how many decimals should be
cut off from the actual resolution of
the current measurement. The fil-
st
ter is a 1
order digital filter.
None
means "No additional filter-
ing" and "Use the modules own resolution" respectively.
Note.
You cannot "worsen" the resolution to less than one significant digit.
In deviation measurement the read-
ing of the main port is compared to
a reference value. The reference
value is subtracted from the read-
ing of the main port.
Start the deviation measurement by
selecting:
Menu
/
D
Window 2 Setup
and
/
B
needed,
or
,
Special Measurements
/
1
Deviation Meas...
.
You can enter the reference value
when you start the special measurement mode.
Note.
Keep in mind that when the displayed deviation reading is small
compared to the actual reading, a significant part of the deviation
value may be measurement error. See the specifications for mea-
surement errors at the actual measurement level.
Special Measurements
S P E C I A L F I L T E R I N G
Q u a n t i t y
P o r t
, if
U n i t
T i m e C o n s t a n t
D e c i m a l s
,
T h e s e t t i n g - 1 , f o r i n s t a n c e , m e a n s o n e
.
l e s s d e c i m a l s .
Q u a n t i t y
P o r t
M e a s u r e d e v i a t i o n f r o m
, if
,
A N D R E S O L U T I O N
s
D E V I A T I O N M E A S U R E M E N T
t h e v a l u e
° C
53