Zeroing A Pressure Module - BEAMEX MC5-IS User Manual

Nstrinsically safe multifunction calibrator
Hide thumbs Also See for MC5-IS:
Table of Contents

Advertisement

Connecting and Removing External Pressure Modules

Zeroing a Pressure Module

Next...
38
Startup and Basic Operation
An external pressure module may be connected and removed at
any time. If a removed module was part of an active measurement,
MC5-IS automatically changes the measurement to a suitable in-
ternal pressure module. MC5-IS also emits a "beep" to inform you
of the fact that the external pressure module used for pressure
measurement was disconnected.
If the selected pressure module does not display zero gauge pres-
sure when the applied pressure is zero, the module has to be ze-
roed.
Open the appropriate window setup menu (
Setup
Window 2 Setup
or
/
B
sure Module
. If a secondary pressure module is active in the se-
lected window, a pop-up menu will appear for choosing either to
zero the primary or the secondary pressure module.
NOTE!
Zeroing a pressure module is especially important when the
operating position of MC5-IS is changed or the location of MC5-
IS is changed in the vertical direction. Both of the above men-
tioned factors affect notably on the pressure measurement
modules. Measuring pressure below 100 mbar (approx. 40 iwc)
should be done with a firmly mounted MC5-IS (e.g. placed on a
table top).
Special Measurements on page 51
Alarm Limit Settings on page 71
Special Generations on page 67
Calibration, see Part D.