Using A Compensation Loop; 2-Wire System - BEAMEX MC5-IS User Manual

Nstrinsically safe multifunction calibrator
Hide thumbs Also See for MC5-IS:
Table of Contents

Advertisement

Using a Compensation Loop

MC5-IS sources current through the resis-
tor and the compensation loop from the
two left side terminals. MC5-IS measures
the voltage drop across the entire current
loop and across the compensation loop.
If the compensation loop and the connec-
tion wires of the resistor are identical, MC5-
IS can compensate for the resistance of
the connection wires.
T/C
INT. RJ
T/C, Low V
T/C
WIRES
ONLY
SENSOR MEASURE & SIMULATE
When the compensation loop wiring sys-
tem is used, MC5-IS displays the text "
wire
".
94
Advanced Operation and Configuration
HART
I
4 w meas
®
meas/sink
R, RTD
R, RTD
V,
I
,
sink
V,
,
Com
meas/sim
3 w meas
Uo: 16 V
Io: 24 mA
Po: 145 mW
OUTPUT
MEASURE
ET
E
Low V

2-wire System

Calibrator sources current through the re-
sistor and measures the voltage drop
across the same terminals.
The result is acceptable, if the resistance
of the connection wires is low.
T/C
INT. RJ
T/C, Low V
T/C
WIRES
ONLY
SENSOR MEASURE & SIMULATE
3-
HART
I
4 w meas
®
meas/sink
R, RTD
R, RTD
V,
I
,
sink
V,
,
Com
meas/sim
3 w meas
Uo: 16 V
Io: 24 mA
Po: 145 mW
OUTPUT
MEASURE
ET
E
Low V

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents