Performing lung function tests
6.4

About posts and bronchodilation

6.5

Performing basic spirometry without gas supply

6.5.1

Performing a complete test

Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
A test and a post are performed together to determine the response to
bronchodilating asthma medication.
After performing an FVC test or an FVL test, the patient is treated with a
bronchodilator. Approximately 10 to 20 minutes after the medication, when
the bronchodilator shows effect, the test is repeated as a post. The results of
the previous test and the post are then compared on the result screen and on
the test protocol. Posts can only be added to an existing test within 24 hours.
Required materials
A new spirette for each patient
For an FVL test, an MVV test, or an SVC test, a nose clip is required.
Prerequisites
You have added the patient data to the database of EasyOne Pro/LAB.
Or the patient data is already in the database of EasyOne Pro/LAB.
You are wearing disposable gloves.
You have instructed the patient on how to perform the test before you
start with the test.
Adding patient data to the database of EasyOne Pro/LAB, 100
About preparation of spirometry and instructions for the patient, 49
WARNING
PATIENT HEALTH HAZARD
Performing spirometry can cause an asthma attack or bronchospasm.
If the patient is on medication, check the contraindications for the medica-
tion.
CAUTION
PATIENT HEALTH HAZARD
Pulmonary function tests require maximum effort on the part of the patient and
may lead to dizziness.
Make sure the patient cannot be injured by objects in the vicinity if the patient
falls.
Watch the patient for signs of dizziness and support the patient if necessary.
Do not leave the patient unattended during a test.
93
51/171