NDD EasyOne Pro Operator's Manual page 73

Hide thumbs Also See for EasyOne Pro:
Table of Contents

Advertisement

Performing lung function tests
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
2 CAUTION! Incorrectly inserted spirette can lead to inaccurate
measurement, false results, and false diagnosis.
Fully insert the spirette into the spirette holder as depicted below, but keep
the partly unwrapped wrapper on the mouthpiece of the spirette.
 The shape of the spirette and the spirette holder guide the orientation
of the spirette. You can only insert the spirette fully if it is in the correct
orientation.
3 On the main menu, choose Patients.
4 Select the patient that you want to test.
5 At the bottom of the screen, choose Test.
6 Choose DLCO.
7 If the environment conditions screen is displayed, verify or adapt the
current values and choose
 The temperature and humidity sensor measures the required data
automatically. However, check these values for plausibility. For example, if
the temperature and humidity sensor is too close to the fan on the back of
EasyOne Pro/LAB, the measured temperature and humidity can be wrong.
 If the patient has already performed a test on the same day, a selection
window is displayed. Proceed with step 8.
 If the patient has not performed a test on the same day, the screen of
the selected test is displayed with a pop-up window that asks you to insert
the DLCO barriette. Proceed with step 9.
8 If the selection window is displayed, do one of the following:
a) To add a trial to the existing test, choose
b) To add a post to the existing test after administering a bronchodilator,
choose
Add
Post.
c) To create a new test that is unrelated to the existing test, choose Test.
 The DLCO test screen is displayed with a pop-up window that asks you
to insert the DLCO barriette.
Confirm
>>.
Add
Trial.
93
73/171

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for NDD EasyOne Pro

This manual is also suitable for:

Easyone pro lab

Table of Contents