Calibration check
11.2.2

Performing and assessing biological quality control

Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
To perform biological quality control, you perform a simple FVC test or FVL test
with your known subject. You can also use biological quality control for the
DLCO test. To be able to compare the measurements of a known subject with
previous measurements, you can use the biological quality control
functionality of EasyOne Pro/LAB.
Prerequisite
You have one or more known subjects in your organization that are
available long-term.
 To perform biological quality control
1 Choose
Utilities > Calibration Check > Biological Quality
2 If you test the subject for the first time, do the following:
a) Choose
Add New BioCal
b) Choose New.
c) Enter the required data and choose OK.
d) From the patient selection list, select the subject that you have just
added.
e) Choose
Add Selected
3 If you test a known subject that you have tested before, select the known
subject from the BioCal subject selection list.
4 Choose
Add New
Test.
5 Choose a test.
6 Perform the test like a regular spirometry or lung function test.
 To assess the biological quality control
1 From the BioCal subject selection list, select the known subject.
2 To view a visualization of the parameters over time, choose Graph.
3 Evaluate the quality criteria for biological quality control.
4 To print a report, choose the reports button.