Connectivity and data exchange
5.2.5.2
Initiating a spirometry test with the direct-start workflow
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
With the direct-start workflow, you initiate spirometry from your EMR system.
All plug-ins support this workflow.
Prerequisite
You have configured any one of the available plug-ins.
About the HL7 plug-in, 43
About the GDT plug-in, 43
About the XML file exchange plug-in, 44
About custom plug-ins, 44
1 Launch your EMR system.
2 Choose the external devices menu.
3 Select either EasyOne Connect or EasyOne Pro/LAB and confirm.
EasyOne Connect is launched automatically.
In EasyOne Connect or on EasyOne Pro/LAB, the screen for entering
patient data is displayed.
4 CAUTION! False diagnosis: The predicted values and the system
interpretation are based on the patient demographic data. Therefore,
wrong patient data can cause false results. Enter patient data
meticulously. Double-check that you have entered the patient data
correctly.
Fill out those fields that are marked with an asterisk (*).
You cannot complete adding patient data without filling out these
fields.
5 In all other fields, enter additional patient data if necessary.
6 Choose OK.
7 Choose the desired test and confirm.
8 Perform spirometry with the patient.
9 If required, print an additional report from the test result screen.
10 Choose Finish.
The test results are sent to your EMR system, and EasyOne Pro/LAB is
closed automatically.
Your EMR system receives the test results.
11 In your EMR system, confirm the received test results and proceed as
required.
Performing lung function tests, 48