Performing lung function tests
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
4 Let the patient take a break.
 The length of the break varies according to the length of the previous
trial. In any case, a countdown for the length of the break is displayed on
the results screen.
 When the countdown is up, a
 To add trials
1 Choose Start.
 Again, preparations are performed automatically.
 Hold the flow sensor away from the patient because gas is flushed from
the patient tube. If the patient inhales the gas that is flushed, this can
influence the later measurement.
 After all preparations are finished, Tidal Breathing is displayed in the
colored status bar.
2 Repeat the previous procedure To perform the MBW breathing maneuver
and this procedure To add trials until the green Session Complete!
message is displayed.
 In the
Test Information
3 Do the following:
a) Review the test quality that is displayed in the
b) Decide whether the test quality is sufficient or not.
c) If necessary, add a trial and repeat the procedure To perform the MBW
breathing maneuver.
 To end the MBW test
1 Grip the spirette with the wrapper around again and pull the spirette out of
the spirette holder. Do not touch the spirette.
2 Dispose of the spirette together with the wrapper.
Start
button is displayed.
area the quality grade for the test is displayed.