6.6.7

Closing the gas valves and disconnecting the gas supply

Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
When not in use, the manufacturer recommends that you close the main valve
on the primary side only and keep the gas tubes connected. This way, you do
not have to find the right pressure level on the secondary side the next time. If
you want to move EasyOne Pro/LAB, you must close both valves, and you can
disconnect the gas tubes.
A
A
Main valve, primary side
C
Manometer, secondary side
E
Port for the DLCO gas tube or the O
 Gas cylinder fittings—same for central gas supply
 To close the gas valves
1 On the primary side, close the main valve A.
 The main valve A closes clockwise.
 The pressure that is displayed on the manometer on the primary side B
does not drop, but stays the same because the pressure is not relieved.
This is correct.
B
B
Manometer, primary side
D
Pressure relief valve, secondary side
gas tube
2
C
D
E
93
84/171