Performing the breathing maneuver for the FVL test

Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
With the flow volume loop test, a deep inhalation directly follows the
exhalation maneuver.
Required material
A nose clip
Prerequisite
The temperature and humidity sensor measures the required data
automatically. However, check these values for plausibility. For example,
if the temperature and humidity sensor is too close to the fan on the back
of EasyOne Pro/LAB, the measured temperature and humidity can be
wrong.
If you choose to adjust the values indicated by the temperature and
humidity sensor, you must enter the environment temperature within 1°C
or 1.8°F accuracy for this test.
1 Tell the patient to breathe at rest.
2 Tell the patient to fill the lungs completely.
3 Tell the patient to take the spirette into the mouth and to seal the lips
around the spirette.
 The patient must not block the opening with the tongue or teeth or bite
down excessively on the spirette.
4 Tell the patient to exhale as hard and as fast as possible and to continue
blowing out until the lungs are completely empty.
5 Tell the patient to keep the spirette in the mouth, to breathe in, and to
continue inhaling until the lungs are completely filled again.
6 Tell the patient to take the spirette out of the mouth and to breathe out
again.
 If the trial has been acceptable, the green quality message is displayed.
 If the trial has not been acceptable, a yellow or red quality message is
displayed suggesting how to improve the breathing maneuver.