Calibration check
11
Calibration check
11.1
Performing a calibration check
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
CAUTION
FALSE DIAGNOSIS
For unforeseen reasons, malfunction of EasyOne Pro/LAB can lead to false re-
sults and false diagnosis.
Perform calibration checks periodically.
Calibration check, 122
Checking for correct operation of EasyOne Pro/LAB, 127
Required materials
A calibration syringe (separately available)
A spirette cal check adapter (separately available)
A spirette
Prerequisite
You have configured the correct syringe volume.
You have configured either Single Flow calibration or Multi Flow
calibration.
Utilities > Configuration > Test > CalCheck, 117
The ultrasonic sensor of EasyOne Pro/LAB does not require calibration. Only
unforeseeable reasons can lead to a malfunction that affects results. In these
cases, you can determine whether EasyOne Pro/LAB operates normally with a
simple calibration check. Therefore, a regular calibration check is
recommended to ensure the reliability of the tests. The American Thoracic
Society (ATS) recommends that calibration must be checked periodically.
If you do not reach ±3.5% accuracy, troubleshoot. If you are not able to
remedy the defect, contact your EasyOne dealer or the ndd Servicing
Department.
1 Unwrap the spirette.
2 Insert the spirette into the flow sensor.
The shape of the spirette and the spirette holder guide the orientation
of the spirette. You can only insert the spirette if it is in the correct
orientation.
3 Attach the spirette cal check adapter to the calibration syringe.
To avoid any temperature differences, fixate the syringe with two
fingers at the one end of the syringe only.
126