NDD EasyOne Pro Operator's Manual page 90

Hide thumbs Also See for EasyOne Pro:
Table of Contents

Advertisement

Performing lung function tests
Code Affecting
Test type
30
DLCO
31
MBW
32
MBW
33
DLCO
34
MBW
35
MBW
46
All
47
All
48
All
50
All
51
All
52
FVC, FVL
 Quality messages for trials
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Description
Allow device to warm up.
CO sensor drift detected.
Let the instrument warm up and repeat the test. If the error reappears, contact
service.
N
concentration at end of test too high – Longer washout required.
2
Tracer concentration at end of test too high. Detailed criteria: tracer concentration at
end of test is >2% N
.
2
Repeat the test, encouraging the patient to continue breathing until at least 3
breaths with concentrations below 2% N
The automatic calibration of the CO
Repeat the test. If the error reappears, contact service.
CO level too high – Please respect waiting time between trials.
CO level during tidal breathing too high.
Wait at least four minutes between DLCO tests. Also check the motor and valve unit. If
the error occurs repeatedly, contact service.
Tidal breathing volume too low.
Tidal breathing volume too low.
Detailed criteria:
Tidal volume outside 4 to 30 mL/kg body weight,
OR outside 0.25 to 0.3 L if the body weight is not stated.
Ask patient to breathe regularly, calmly and within the suggested range.
Check patient technique, gas supply and check for leaks.
Volume drift during washout test detected. The volume drift correction could not be
performed and/or the parameters IRV and/or ERV are smaller than zero.
Check for correct test performance and/or for leaks. If the error occurs repeatedly,
contact service.
Timeout identified.
A timeout has occurred during the maneuver.
Follow all test instructions.
Device error.
An unexpected device error occurred.
Check if sensor is connected correctly and repeat trial.
Invalid trial, please repeat trial.
Trial did not meet acceptability criteria.
Repeat trial.
No error or warning, test is OK.
Error (51): Test started too early.
Early-start detected.
Wait until the instrument signals to start the test.
Error (52): Unable to detect maneuver.
Unable to detect maneuver.
Retry maneuver and follow the instructions that are shown during the maneuver.
are achieved.
2
sensor failed.
2