NDD EasyOne Pro Operator's Manual page 11

Hide thumbs Also See for EasyOne Pro:
Table of Contents

Advertisement

Safety information
Electric shock or
hot surfaces
Cross-
contamination
Patient health
hazard and device
malfunction
False diagnosis
False diagnosis
Patient health
hazard and
false diagnosis
False diagnosis
Failed update and
data loss
Malfunction
Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
Do not unscrew and open EasyOne Pro/LAB or any parts of it.
Tuberculosis or other diseases spread by droplet nuclei and can cause aerosol
infection.
Proper attention to environmental engineering controls, such as ventilation,
air filtration or ultraviolet decontamination of air, must be used.
For spirometry tests with gas supply, off-specification test gas can cause ad-
verse reaction, poisoning, allergic reaction, or other harm to the patient.
Also, off-specification test gas can cause device malfunction.
Use only appropriate gas, according to the test gas requirements.
Description of test gas requirements, 136
For unforeseen reasons, malfunction of EasyOne Pro/LAB can lead to false re-
sults and false diagnosis.
Perform calibration checks periodically.
Calibration check, 122
Checking for correct operation of EasyOne Pro/LAB, 127
EasyOne Pro/LAB can be damaged when dropped or during transport.
Do an incoming inspection when you first receive EasyOne Pro/LAB. If you de-
tect a damage, contact your EasyOne dealer or the ndd Servicing Department.
Do not drop EasyOne Pro/LAB.
If EasyOne Pro/LAB has been dropped, check for correct operation of
EasyOne Pro/LAB.
Checking for correct operation of EasyOne Pro/LAB, 127
Inadequate qualification of the personnel can put the patient's health at risk
and can result in incorrect execution of tests, false results, and false diagnosis
and/or high costs.
EasyOne Pro/LAB must only be used by qualified personnel.
Intended audience of this Operator's Manual, 3
Setup with wrong environment data can cause false results and false diagno-
sis.
Verify the set up environment data.
Description of environment configurations, 119
An update may fail.
For various other reasons, data loss can occur and is generally unforeseen.
Perform backups of the database of EasyOne Pro/LAB frequently.
About exporting comprehensive data from EasyOne Pro/LAB, 104
Non-original accessories, like tubes or cables can cause malfunction.
Only use original accessories by ndd.
List of order numbers and accessories for EasyOne Pro/LAB, 136
14
11/171