About Using The Touchscreen - NDD EasyOne Pro Operator's Manual

Hide thumbs Also See for EasyOne Pro:
Table of Contents

Advertisement

Introduction
3.2.5

About using the touchscreen

Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
 To turn EasyOne Pro/LAB off
1 If you have been performing lung function tests with gas supply before,
close the gas valves.
 You can keep the gas supply connected.
2 On the main menu, choose
 The shut down window is displayed.
3 From the shut down window, choose
 The pressure is released from the gas tube.
4 After the pressure has been released, choose OK.
 EasyOne Pro/LAB shuts down.
5 Optionally, after EasyOne Pro/LAB has shut down, you can turn off the
mains switch on the back.
 Closing the gas valves and disconnecting the gas supply, 84
Single-touch control
The touchscreen supports single-touch. That means you can use your finger
as a pointer and for single taps, which is comparable to a mouse pointer with
single clicks.
You only need to tap any element once. There is no need for double taps.
Using the touchscreen with gloves
You can use the touchscreen wearing disposable gloves made of any
material. Standard disposable gloves used in hospitals and general
practioners' offices work fine with the touchscreen.
Shut
down.
Pressure
Release.