List Of Quality Criteria For Biological Quality Control; About A Dlco Calibration Check; About An Mbw Calibration Check - NDD EasyOne Pro Operator's Manual

Hide thumbs Also See for EasyOne Pro:
Table of Contents

Advertisement

Calibration check
11.2.3

List of quality criteria for biological quality control

Quality criterion
Limitation
BA
BB
1 value outside ±2 SD
BC
1 value outside ±3 SD
BD
≥ 4 values outside ±1 SD
≥ 2 values outside ±2 SD
> 1 value outside ±3 SD
≥ 10 values, consecutive values on same side of mean
BF
 Specifications of quality criteria for biological quality control
11.2.4

About a DLCO calibration check

11.2.5

About an MBW calibration check

Operator's Manual V5.0 • EasyOne Pro/LAB Rev. 05
© ndd Medizintechnik AG • Technoparkstrasse 1, 8005 Zurich, Switzerland • www.ndd.ch
For each parameter (FEV
1
criterion is calculated. Additionally the lowest grade of all parameters is
reported.
A DLCO calibration check can be performed by using biological quality
control.
The American Thoracic Society (ATS) recommends checking DLCO calibration
on a weekly basis or whenever you suspect irregularities.
 Operator's Manual of your DLCO simulator
Performing and assessing biological quality control, 125
Performing lung function tests with gas supply, 65
About the trend view with EasyOne Pro/LAB, 102
The American Thoracic Society (ATS) recommends that testing of biological
controls should be performed at least monthly.
 Performing a calibration check, 122
Performing lung function tests with gas supply, 65
About the trend view with EasyOne Pro/LAB, 102
, FVC, FEV
, DLCO, VI, VA) an individual quality
6
Additional limitation
only consecutive values on same
side of mean
only consecutive values on same
side of mean
126
126/171